Page 90 of 108
Is mounted
with 4 screws
Mounting:
Massage motor on a plate
Mounting of the screw with max. torque 2 Nm
Mounting of massage motor by using brackets:
Screw
Slat
Bracket
Massage
motor
Nut
The massage unit is mounted with 4 x M6 roundheaded machine screws with flat underside. 15 to 20 mm long + the thickness of the bracket.
Torque max. 2-3 Nm.
2 brackets must be used - one on each side of the slat.
6. MJB (MEDLINE
®
CARELINE
®
)
The MJB (Modular Junction Box) is designed for use together with the CB20 and other OpenBus
TM
control boxes.
The MJB makes it possible to connect handsets, attendant controls and accessories with the CB20/CB6S/CB16. The
CB20 has 2 modular jack interfaces, but by using the MJB the connection possiblities are greatly increased. One
“modular jack” channel is used for connecting to the control box.
7. SLS (MEDLINE
®
CARELINE
®
TECHLINE
®
)
LINAK has designed a switch that can be mounted in connection with the 24V DC actuators: LA12, LA22, LA28,
LA28S, LA31, LA32, or LA34 and a control box on e.g. a bed frame.
The Safety Limit Switch (SLS). The SLS Switch is primarily used as a Limit Switch in systems consisting of a control
box, LA28/28S and/or LA32.
As an example the SLS can be installed on the actuator cable where it disconnects the actuator in either inward or
outward direction when activated. The SLS can also be used as a signal control directly connected to a control box.
8. SMPS30 (MEDLINE
®
CARELINE
®
)
The SMPS30 is a very powerful Switch Mode Power Supply typically used
for Treatment or Examination Couches. The SMPS30 is an eco-friendly
solution due to a low standby power consumption compared to tradi-
tional transformer solutions. The universal input voltage makes the SMPS
adaptable to the worldwide market irrespective of the input voltage.
Usage:
• Compatibility: LA43 IC, LA44 IC
• Duty Cycle*: 10%, 2 minutes continuous use followed by 18 minutes not in use.
• Operating temperature: 5°C to 40°C
• Storage temperature: -10°C to 50°C
• Relative humidity: 20% to 90% @ 30°C – not condensing
• Atmospheric pressure: 700 to 1060 hPa
• Altitude: max. 2000 metres above sea level
• Approvals: IEC60601-1:2005 3
rd
ed., ANSI / AAMI ES60601-1:2005, 3
rd
edition, CAN / CSA-22.2 No 60601-1:2008
* AT 10% DUTY CYCLE, MAX. OUTPUT POWER IS POSSIBLE AT AN AMBIENT TEMPERATURE OF 25°C
...to be continued