The ACM (Mini Attendant Control) box is fitted to hospital and care beds for use where patient positioning
must be carefully controlled by the medical staff.
The compact design and simple operation makes it easy for the nursing staff to retain direct control over
critical positioning functions whilst giving the patient a limited degree of adjustment.
The ACM must be mounted correctly on a flat surface to ensure IP degree. A short circuit in the cable can
cause movement. To possibly avoid this risk, choose a OpenBus system.
4. ACM (MEDLINE
®
CARELINE
®
)
6. ACP (MEDLINE
®
CARELINE
®
)
The ACP (Attendant Control Panel) controller is fitted to advanced
hospital and patient care beds for use where patient positioning must
be carefully controlled by medical staff.
The ACP allows nursing staff to retain direct control over critical
positioning functions while giving the patient a limited degree of
adjustment.
It is recommended to clean all surfaces, also surfaces covered by lids and the inside of the lids.
(In the event that the ACP is cleaned in wash tunnels; be aware that surfaces covered by lids and the inside of lids will not be cleaned)
7. DPH Medical (MEDLINE
®
CARELINE
®
)
The DPH is a small two button panel for adjustment of different functions. The DPH fits into a MJB with a modular plug and thereby compatible with
OpenBus.
Usage:
• Ambient temperature: +5° to +40°
• DPH is compatible with the OpenBox control boxes via Modular Junction Boxes MJB5061101-00 as follows:
Control Boxes: CB6S OBL, CB6S OBM, CB6S OBF, CB16 OBL, CB16 OBF, CB20
Modular Junction Box: MJB5061101-00
• DPH1K10-210007 combined with MJB5061101-00 creates the OpenBus codes:
Up Arrow: H0
Down Arrow: H1
• Wrong mounting is not an issue with the MJB5061101-00 and the modular jack plug of the DPH cable. The plug will only fit into the correct ports
of the MJB.
The ACO (Attendant Control OpenBus) is a cost optimised and compact unit with up to 15 buttons that can be
used as Handset keys or lock-outs.
The lock-out function can be made visable by using yellow LED´s.
The ACO is compatible with control boxes that use an OpenBus™ interface i.e. CB6S, CB16 and CB20.
5. ACO (MEDLINE
®
CARELINE
®
)
Usage:
• Compatible with OpenBus™ e.g. CB20, CB6S and CB16.
• Ambient temperature +5˚ to + 40˚C
• For approvals information see Lintra/Approvals/Certificates
• ACO IPX6-Washable is approved according to EN60601-2-52.
In order to comply with the norm the ACO must hang vertically from it’s hook during the washing process.
Recommendations:
• To switch between locked and unlocked position a small knob between the two pushbuttons has to be turned 20° by use of a special key.
The key is for the use of the nursing staff only, there are two types, one is made of plastic the other metal.
• The key has to be ordered separately. Article no. for the plastic key is: 00914516, and the metal key number is: 00914721
• For all types: Attention should be given to ensure that the channels shown correspond to the channels available on the chosen control box
• Always use Locking ring and cables with O-rings.
• Locking ring and cables with O-rings must be fitted to ensure IP degree.
• If other front covers than standards are requested, the customer must design them..